Hanna GroLine Nutrient Dosing System

In stock
SKU: HI981413-10
Regular price $530.00

Hanna®

GroLine Nutrient Dosing System

The HI981413-10 Hanna Nutrient Dosing System is engineered for maintaining the concentration of fertilizer in the nutrient solution used in hydroponics and irrigation water.

The nutrient concentration in the water used for irrigation is critical for the successful propagation and growth of plants.

Simply insert the probe and injection valve in-line with the recirculation pump and provide the chemical to be dosed.

  • Includes Hanna HI30033 EC/TDS/TEMP Probe.
  • Injection Valve Fitting & Tubing Kit Included.
  • Programmable alarm system and multicolored LCD display.
  • Perfect for small hydroponics systems.

    GroLine EC/Temperature Probe

    The HI981413 uses the HI30033 amplified probe that incorporates both EC (TDS) and temperature sensors and connects to the controller with a single waterproof Quick Connect DIN connector.
    • PVDF body
    • 1/2” threaded fitting for insertion to an in-line “T” fitting or the flow cell

    Automatic Proportional Pump Control

    The peristaltic dosing pump can be controlled by simple on/off or more advanced proportional control which helps prevents overshooting of the set point.

    Adjustable Flow Rate The dosing pump flow rate is adjustable from 0.5 to 3.5L/h, allowing for better control in maintaining a desired set point.

    Acid Tank Level/Flow Switch Input

    A connection to an optional level controller or flow switch. This input can be used to disable the dosing pump when there is no chemical left in the reservoir tank or there is no flow due to the pump being turned off.

    Multicolored LCD Display

    A multi-colored LCD provides a quick way to see the status of the pool controller. The display will show green when operation normally and red while in an alarm state.

    Quick Connect Probe Input

    The Quick Connect DIN connector creates a waterproof seal with the meter making it ideal for reducing electrical noise issues with the connection caused by humid environments.





     

    Specification Name Detail
    SKU HI981413
    Temperature Range -5.0 to 105°C (23.0 to 221.0°F)
    Temperature Resolution 0.1°C (0.1°F)
    Temperature Accuracy (@25°C/77°F) ±0.5°C (±0.9°F)
    EC Range 0.00 to 10.00 mS/cm
    EC Resolution 0.01 mS/cm
    EC Accuracy (@25°C/77°F) ±2% F.S.
    TDS Range 0 to 4500 ppm (TDS conversion factor 0.45)
    0 to 9900 ppm (TDS conversion factor 0.99)
    TDS Resolution 1 ppm
    TDS Accuracy (@25°C/77°F) ±2% F.S.
    TDS Conversion Factor TDS conversion factor selectable from 0.45 to 0.99
    Temperature Compensation automatic
    Temperature Coefficient can be set from 0%/°C to 2.4%/°C; default is 1.9%/°C
    Calibration EC: user calibration: automatic, one-point with standard solution (1.413 or 5.000 mS/cm)
    Process calibration: single point, adjustable (±0.50 mS/cm around measured value)
    TDS: through EC calibration
    Pump Control selectable flow rate (0.5 to 3.5 L / hour; 0.13 to 0.92 G/hour)
    manual control for pump priming
    Alarms high and low with enable / disable option
    triggered after 5 sec. if controller records a set of consecutive readings over / under threshold values level with enable / disable option
    overtime protection (1 to 180 min. or off)
    intuitive alarm system using red, light green, and green color-coded backlight
    External Event Input input for level controller or flow switch to disable dosing pump in the event of no chemical when using a level controller or no flow when using a flow switch - galvanically isolated
    Alarm Relay Output SPDT 2.5A / 230 VAC
    Power Supply 100—240 VAC, 50/60 Hz
    Power Consumption 15 VA
    Casing wall mounted, built-in pump, IP65 rated
    Environment 0-50°C (32-122°F), max. 95% RH non-condensing
    Dimensions 90 x 142 x 80 mm (3.5 x 5.6 x 1.8”)
    Weight 910 g (32 oz.)
    Ordering Information HI981413-00 is supplied with HI30033 EC/TDS/temperature probe, conductivity calibration solution, 20 mL (3), power connection cable, instruction manual and quality certificates for instrument and probe.
    HI981413-10 (with in-line mounting kit) is supplied with HI981413 controller, HI30033 EC/TDS/temperature probe, controller aspiration filter, controller injector, 1/2” thread, saddle for Ø 50 mm pipe (2), aspiration PVC tube (flexible) (5 m), dispensing PE injection tube (rigid) (5 m), valves (2), conductivity calibration solution, 20 mL (3), power connection cable, instruction manual and quality certificates for instrument and probe.
    HI981413-20 (with flow-cell mounting kit) is supplied with HI981413 controller, HI30033 pH/temperature probe, flow cell for HI981413/BL101, mounting panel assembly for HI981413/BL101, controller aspiration filter, controller injector, 1/2” thread, saddle for Ø 50 mm pipe (3), aspiration PVC tube (flexible) (5 m), dispensing PE injection tube (rigid) (15 m), tubing adapter 1/2” - 6 mm with racord (2), valves (2), conductivity calibration solution, 20 mL (3), power connection cable, instruction manual and quality certificates for instrument and probe.
    Recommended Probe HI30033 EC/TDS/Temperature probe with quick connect DIN connector - galvanically isolated